Auto change over for medical gas pipeline consist of Oxygen &
Nitrous Oxide gas control with manifold system.
Oxygenx medical gas manifold & control is a pressure reducer
semi/ automatic change over system, which has been specially
designed for the distribution of compressed medical gases from
either bank of cylinders to medical gas pipeline system. It
ensures uninterrupted gas supply to distribution network when
connected to high pressure (140 bar / 2030 psi).
Technical Data Control Panel
- Upstream Pressure – 140 Bar (2030 psi) max
- Delivery Pressure – 4 Bar (58 psi)
- Flow rate at 4 Bar
-
Oxygen – 660 lpm / 1500 lpm max
- Nitrous Oxide – 710 lpm max
- Working temperature – 0-50 degree centigrade
-
Pressure of Safety Valve – 140>10 bar = 10.5 bar / 180
psi
10> 5 bar = 7.5 bar /108 psi
-
Bank Switching – Semi Automatic Manual – Automatic
(Pneumatic Switching)
- Electrical – 230 Volt AC / 12 Volt DC
External adapter
for audio visual alarm
- Manifold Configuration – 2,4 & 6 cylinders (module) can be
interconnected to expand bank size
Modular manifold has been designed to provide user flexibility &
can be configured to any desire combination. Manifold headers
consist of high pressure pipe with gas specific tail pipe
connected to Non Return Valve (NRV)
The audio visual alarm is provided to monitor the operational
status besides analog pressure gauge/ digital pressure gauge on
the panel.